Calibration of multiple LDR brachytherapy sources
Larry A DeWerd1, John A Micka, Shannon M Holmes
1Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 55106, USA. ladewerd@wisc.edu
Medical physicists need new methods to verify the strength of prepackaged brachytherapy sources. This study evaluates measurement techniques for various low dose rate brachytherapy source geometries, offering practical solutions for accurate verification.
Area of Science:
- Medical Physics
- Radiation Oncology
- Nuclear Medicine
Background:
- The increasing use of prepackaged low dose rate brachytherapy sources necessitates reliable verification methods.
- Medical physicists are responsible for ensuring accurate source strength before patient treatment.
Purpose of the Study:
- To experimentally evaluate the impact of different multiseed geometries on source strength measurements.
- To develop practical verification methods for prepackaged brachytherapy sources.
Main Methods:
- Review of existing guidelines for prepackaged brachytherapy sources.
- Experimental measurement of multiseed strands, coils, and sterile seed cartridge packs using a well-type ionization chamber.
- Evaluation of measurement techniques for sources within needles, including removal into sterile inserts.
Main Results:
- Multiseed strands and coils (125I, 103Pd, 192Ir) can be accurately measured in chambers with long, uniform sensitive lengths.
- Sterile seed cartridge packs require a correction factor for accurate measurement.
- Sources within needles cannot be measured directly due to variations in needle composition and thickness; removal into sterile inserts is a viable alternative.
Conclusions:
- Practical methods for verifying the source strength of various prepackaged brachytherapy sources have been established.
- Correction factors are provided where necessary for accurate measurements.
- Maintaining sterility during measurement is crucial, especially when handling multiple seeds.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...