Related Experiment Video
Updated: Jul 18, 2026

Ex Vivo Porcine Experimental Model for Studying and Teaching Lung Mechanics
Published on: April 19, 2024
Use of dynamic compliance for open lung positive end-expiratory pressure titration in an experimental study
Fernando Suarez-Sipmann1, Stephan H Böhm, Gerardo Tusman
1Intensive Care Unit, Fundación Jiménez Díaz-Capio, Avenida de los Reyes Católicos 2, 28010 Madrid, Spain. fsuarez@fjd.es
Objective:
We tested whether the continuous monitoring of dynamic compliance could become a useful bedside tool for detecting the beginning of collapse of a fully recruited lung.
Design:
Prospective laboratory animal investigation.
Setting:
Clinical physiology research laboratory, University of Uppsala, Sweden.
Subjects:
Eight pigs submitted to repeated lung lavages.
Interventions:
Lung recruitment maneuver, the effect of which was confirmed by predefined oxygenation, lung mechanics, and computed tomography scan criteria, was followed by a positive end-expiratory pressure (PEEP) reduction trial in a volume control mode with a tidal volume of 6 mL/kg. Every 10 mins, PEEP was reduced in steps of 2 cm H2O starting from 24 cm H2O. During PEEP reduction, lung collapse was defined by the maximum dynamic compliance value after which a first measurable decrease occurred. Open lung PEEP according to dynamic compliance was then defined as the level of PEEP before the point of collapse. This value was compared with oxygenation (Pao2) and CT scans.
Measurements And Main Results:
Pao2 and dynamic compliance were monitored continuously, whereas computed tomography scans were obtained at the end of each pressure step. Collapse defined by dynamic compliance occurred at a PEEP of 14 cm H2O. This level coincided with the oxygenation-based collapse point when also shunt started to increase and occurred one step before the percentage of nonaerated tissue on the computed tomography exceeded 5%. Open lung PEEP was thus at 16 cm H2O, the level at which oxygenation and computed tomography scan confirmed a fully open, not yet collapsed lung condition.
Conclusions:
In this experimental model, the continuous monitoring of dynamic compliance identified the beginning of collapse after lung recruitment. These findings were confirmed by oxygenation and computed tomography scans. This method might become a valuable bedside tool for identifying the level of PEEP that prevents end-expiratory collapse.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Respiratory Volumes and Capacities I