Related Experiment Video
Updated: Jul 18, 2026

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Quantitative evaluation of three calibration methods for 3-D freehand ultrasound
François Rousseau1, Pierre Hellier, Marloes M J Letteboer
1IRISA, CNRS-INRIA-INSERM-University of Rennes 1, F-35042 Rennes, France. rousseau@lsiit.u-strasbg.fr
Abstract:
In this paper, three different calibration methods for three-dimensional (3-D) freehand ultrasound (US) are evaluated. Calibration is the process of estimating the rigid transformation from US image coordinates to the coordinate system of the tracking sensor mounted onto the probe. Calibration accuracy has an important impact on quantitative studies. Geometrical precision can also be crucial in many interventions and surgery. The proposed evaluation framework relies on a single point phantom and a 3-D US phantom which mimics the US characteristics of human liver. Four quality measures are used: 3-D point localization criterion, distance and volume measurements, and shape based criterion. Results show that during the acquisition procedure, volumetric measurements and shapes of the reconstructed object depend on probe motion used, particularly fan motions for which errors are larger. It is also shown that accurate calibration is essential to obtain reliable quantitative information.