Ten-year follow-up of 360 degrees intrastromal corneal rings for myopia
Related Experiment Video
Updated: Jul 18, 2026

Correction of Presbyopia by Monocular Bi-Aspheric Ablation Profile
Published on: September 20, 2024
Andrew R Schwartz1, Bernard O Tinio, Fatema Esmail
1Cornea Service, Department of Ophthalmology, Mount Sinai School of Medicine, New York, NY , USA.
Purpose:
To determine the safety, efficacy, and stability of intrastromal corneal rings 10 years after placement for myopia.
Methods:
Ten eyes with myopia treated with the placement of 360 degrees complete intrastromal corneal rings with the ends sutured together were evaluated at 10-year follow-up for the following: uncorrected visual acuity (UCVA), best spectacle-corrected visual acuity (BSCVA), manifest refraction spherical equivalent (MRSE), induced manifest refractive cylinder, keratometry readings, slit-lamp findings, and ultrasound central pachymetry. Ten-year data were compared to 1-year results to assess visual stability over time.
Results:
No statistically significant difference was noted between UCVA at 1 year and 10 years. At the 10-year examination, 90% of patients had BSCVA > or = 20/25 and 100% had BSCVA > or = 20/30. There was no statistically significant difference between MRSE at 1 year and 10 years. At the 10-year examination, induced manifest refractive cylinder was >1.00 diopter (D) in 0% of eyes and < or = 0.25 D in 60% of eyes. There was no statistically significant difference between mean central keratometric power at 1-year follow-up compared with 10-year follow-up. No statistically significant difference was noted in central corneal thickness between 1 and 10 years in the eyes studied.
Conclusions:
Intrastromal corneal rings are an effective and stable method of correcting mild myopia based on optic parameters.