Late re-operation for aortic and mitral Starr-Edwards ball valve prostheses
Shigeaki Aoyagi1, Shuji Fukunaga, Koichi Arinaga
1Department of Surgery, Kurume University School of Medicine, 67 Asahi-machi, Kurume 830-0011, Japan. aoyagi@med.kurume-u.ac.jp
Cloth wear is a significant issue in Starr-Edwards cloth-covered ball valves implanted for over 15 years, potentially causing hemolysis and requiring replacement. Non-cloth-covered valves showed minimal wear, indicating cloth wear is a primary concern for long-term function.
Area of Science:
- Cardiovascular Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- Starr-Edwards ball valves have been used for decades in cardiac valve replacement surgery.
- Long-term explantation studies are crucial for understanding device durability and failure modes.
- Cloth coverings were introduced to improve tissue ingrowth and reduce paravalvular leaks.
Purpose of the Study:
- To investigate the long-term explanted Starr-Edwards ball valves (≥15 years post-implantation).
- To evaluate the macroscopic condition and identify failure mechanisms, particularly cloth wear.
- To correlate explantation findings with clinical indicators of valve dysfunction.
Main Methods:
- Retrospective macroscopic investigation of explanted Starr-Edwards ball valves (≥15 years).
- Analysis of valve models (cloth-covered vs. non-cloth-covered), position (aortic/mitral), and explantation time.
- Correlation of explant findings with patient re-operation data and clinical signs of hemolysis.
Main Results:
- Cloth wear was observed in all explanted cloth-covered valves, irrespective of position, leading to significant hemolysis.
- Non-cloth-covered valves exhibited only mild ball variance.
- Autologous tissue ingrowth was noted on components, but thrombus formation was absent.
Conclusions:
- Cloth wear is the primary long-term complication of Starr-Edwards cloth-covered ball valves.
- Clinical signs like hemolysis and regurgitation indicate cloth wear and necessitate valve replacement.
- Consideration of cloth wear is essential for patients with these valves >15 years post-implantation.