Related Experiment Video
Updated: Jul 18, 2026

Drug-Induced Sleep Endoscopy (DISE) with Target Controlled Infusion (TCI) and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Comparison of automatic and continuous positive airway pressure in a night-by-night analysis: a randomized, crossover
Wolfgang Galetke1, Norbert Anduleit, Kerstin Richter
1Bethanien Hospital Solingen, Clinic for Pneumology and Allergology, Center for Sleep Medicine and Respiratory Care, Institute for Pneumology, University Witten/Herdecke, Solingen, Germany. galetke@klinik-bethanien.de
Background:
Long-term compliance is suboptimal in the treatment of the obstructive sleep apnea syndrome (OSAS).
Objectives:
We compared the efficacy of and the adherence to automatic continuous positive airway pressure (APAP) and constant continuous positive airway pressure (CPAP) based on a night-by-night analysis.
Methods:
We performed a randomized, single-blind crossover study in 20 patients with moderate-to-severe OSAS. After diagnostic polysomnography and manual titration, patients were treated for 8 weeks with both constant CPAP and APAP in random order. Compliance and leakage were analyzed night by night using the software LOGSoft of the Magellan iPAP device.
Results:
The reduction in the apnea/hypopnea index (baseline 32.9 +/- 19.1/h, CPAP 4.6 +/- 2.9/h, APAP 5.6 +/- 3.6/h; p < 0.001 compared to baseline) and the Epworth Sleepiness Scale (baseline 10.3 +/- 5.7, CPAP 6.6 +/- 4.8, APAP 4.9 +/- 4.6; p < 0.001 compared to baseline) did not significantly differ between the treatment modes. Leakage time and compliance per night were not statistically different (leakage CPAP 31 +/- 57 min, APAP 25 +/- 49 min; compliance CPAP 383 +/- 116 min, APAP 382 +/- 107 min). There was no correlation between leakage and compliance. Thirteen patients (65%) preferred APAP at the end of the study.
Conclusions:
Treatment efficacy and adherence are similar with CPAP and APAP. There is a trend towards lower leakage with APAP therapy. Patients prefer the automatic mode to fixed pressure.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Sleep Apnea
The condition is more prevalent among...
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Blind Procedures