Determination of ultrafilterable prolactin: elimination of macroprolactin interference with a monomeric

Adam M Quinn1, Tara C Rubinas, C Jean Garbincius

  • 1Department of Pathology, Loyola University Medical Center, Maywood, Ill, USA. adamquinn1@gmail.com

Abstract

Insights

Ultrafiltration effectively removes macroprolactin (macroPRL) interference in prolactin (PRL) immunoassays. This method provides a practical alternative for accurately measuring monomeric PRL levels.

Area of Science:

  • Clinical Chemistry
  • Endocrinology
  • Laboratory Medicine

Background:

  • Macroprolactin (macroPRL) causes apparent hyperprolactinemia in up to 25% of elevated prolactin tests.
  • This can lead to unnecessary clinical investigations and diagnostic procedures.

Purpose of the Study:

  • To develop an ultrafiltration method to eliminate macroPRL interference in prolactin (PRL) immunoassays.
  • To establish a practical and precise method for measuring monomeric PRL.

Main Methods:

  • Serum ultrafiltration using a Centricon-100 filter device.
  • Assay of prolactin (PRL) in the serum ultrafiltrate.

Main Results:

  • Ultrafiltrates removed macroPRL, retaining 85% of total PRL immunoreactivity.
  • Fractional recovery of ultrafilterable PRL (uPRL) was 0.78.
  • The uPRL concentration in healthy females was 8.0 +/- 3.1 ng/mL.

Conclusions:

  • Ultrafiltration is a rapid, simple method to eliminate macroPRL analytical interference.
  • Ultrafiltrates are suitable for most PRL immunoassays.
  • This method is a practical alternative to gel filtration chromatography for monomeric PRL estimation.