Summary
The Exeter stem shows excellent mid-term results with no aseptic loosening revisions. The Endurance stem had poor early results, with 20.5% of patients requiring revision for aseptic loosening.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
Background:
- Cemented femoral stems are widely used in hip arthroplasty.
- Comparing different cemented stem designs is crucial for optimizing long-term outcomes.
Purpose of the Study:
- To compare the mid-term clinical performance of two distinct cemented femoral stem designs: the Exeter stem and the Endurance stem.
- To evaluate the incidence of aseptic loosening and the need for revision surgery for each stem type.
Main Methods:
- Prospective study of 118 Exeter stems and 34 Endurance stems implanted by a single surgeon.
- Utilized third-generation cement techniques via a lateral approach.
- Follow-up averaged 7.25 years for Exeter and 4.57 years for Endurance.
Main Results:
- The Exeter stem group experienced 0% revisions for aseptic loosening.
- The Endurance stem group had a 20.5% revision rate for aseptic loosening, with an average failure time of 3.72 years.
- Significant difference in aseptic loosening rates between the two stem designs.
Conclusions:
- The Exeter stem demonstrates excellent mid-term results, supporting its continued use in hip arthroplasty.
- The Endurance stem's poor early performance and high revision rate for aseptic loosening led to its discontinuation.
- Stem design and surface characteristics significantly impact the longevity of cemented femoral components.