Psychophysical evaluation of calibration curve for diagnostic LCD monitor
Masanobu Uemura1, Yoshiyuki Asai, Michihiro Yamaguchi
1Department of Central Radiology, Kinki University Hospital, 377-2 Ohnohigashi, Osaka-Sayama 589-8511, Japan. uemura@radt.med.kindai.ac.jp
Purpose:
In 1998, Digital Imaging Communications in Medicine (DICOM) proposed a calibration tool, the grayscale standard display function (GSDF), to obtain output consistency of radiographs. To our knowledge, there have been no previous reports of investigating the relation between perceptual linearity and detectability on a calibration curve.
Materials And Methods:
To determine a suitable calibration curve for diagnostic liquid crystal display (LCD) monitors, the GSDF and Commission Internationale de l'Eclairage (CIE) curves were compared using psychophysical gradient delta and receiver operating characteristic (ROC) analysis for clinical images.
Results:
We succeeded in expressing visually recognized contrast directly using delta instead of the just noticeable difference (JND) index of the DICOM standard. As a result, we found that the visually recognized contrast at low luminance areas on the LCD monitor calibrated by the CIE curve is higher than that calibrated by the GSDF curve. On the ROC analysis, there was no significant difference in tumor detectability between GSDF and CIE curves for clinical thoracic images. However, the area parameter Az of the CIE curve is superior to that of the GSDF curve. The detectability of tumor shadows in the thoracic region on clinical images using the CIE curve was superior to that using the GSDF curve owing to the high absolute value of delta in the low luminance range.
Conclusion:
We conclude that the CIE curve is the most suitable tool for calibrating diagnostic LCD monitors, rather than the GSDF curve.
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Calibration Curves: Correlation Coefficient
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Receiver Operating Characteristic Plot