Related Experiment Video
Updated: Jul 18, 2026

A Simple Method for Automated Solid Phase Extraction of Water Samples for Immunological Analysis of Small Pollutants
Published on: January 1, 2016
Sample clean-up with sol-gel enzyme and immunoaffinity columns for the determination of bisphenol A in human urine
Kerstin Schöringhumer1, Margit Cichna-Markl
1Department of Analytical and Food Chemistry, University of Vienna, Währinger Strasse 38, A-1090 Vienna, Austria.
Abstract:
The paper describes the development of a simple and highly selective analytical method for the determination of free and total bisphenol A in urine samples. Free bisphenol A levels can be determined after sample clean-up using sol-gel immunoaffinity columns containing anti-bisphenol A antibodies. In determining total bisphenol A levels, the sample pre-treatment procedure consists of sample preparation using an on-line combination of two sol-gel columns, an enzyme column containing glucuronidase and arylsulfatase, and an immunoaffinity column. Bisphenol A can then be quantified by high-performance liquid chromatography and fluorescence detection. The mean recovery was found to be 78% with a standard deviation of 3.4%, the LOD (S/N=3) was 0.2 ng/ml. The method was applied to determine free and total urinary BPA levels of healthy adults and dialysis patients.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018