Oximetry feedback flow control simulation for oxygen therapy
Related Experiment Video
Updated: Jul 17, 2026

Quantitative and Temporal Control of Oxygen Microenvironment at the Single Islet Level
Published on: November 17, 2013
Mario G Iobbi1, Anita K Simonds, Robert J Dickinson
1Department of Bioengineering, South Kensington Campus, Imperial College London, London, SW7 2AZ, UK.
Objectives:
For many with Chronic Obstructive Pulmonary Disease (COPD), arterial oxygen saturation while receiving Long-Term Oxygen Therapy (LTOT) falls below an acceptable threshold (SpO(2) < 90%) for extended periods during routine daily activities. Using a closed-loop controller, we have evaluated a simulated method to automatically regulate the oxygen flow-rate in response to the measured oxygen demand.
Methods:
The closed-loop control scheme was implemented in a computer simulation on Simulink. Feedback from a pulse oximeter was used to maintain a target SpO(2) of 91% by changing the oxygen flow-rate to the patient. The controller was evaluated using a model to approximate the patient's arterial oxygen saturation response, including hypoxic events from artificial disturbances as well as recorded patient oximetry data.
Results:
The simulated controller produced improvement in arterial oxygen saturation throughout a wide range of disturbance frequencies. It suppressed disturbances with periods greater than a couple of minutes by more than -10 dB. When evaluated with patient oximetry recordings, the controller on average reduced the time spent with arterial blood saturation below threshold by 76%. Given the same volume of oxygen, the closed-loop controller also produced a 63% improvement compared to fixed flow-rate LTOT.
Conclusions:
The simulation findings indicate an optimized matching between oxygen supply and demand, maintaining SpO(2) above threshold to improve therapeutic efficacy compared to standard LTOT.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...