Clinical evaluation of applanation resonance tonometry: a comparison with Goldmann applanation tonometry
Per Hallberg1, Anders Eklund, Tomas Bäcklund
1Department of Biomedical Engineering and Informatics, University Hospital of Umeå, S-901 85, Umeå, Sweden.
Purpose:
The purpose of this study was to calibrate and evaluate the precision of the new applanation resonance tonometry (ART) in a clinical study designed in accordance with the International Standard Organization's requirements.
Methods:
This was a prospective, randomized, single-center study, where healthy volunteers and patients participated. A total of 153 eyes were divided into 3 groups with respect to their intraocular pressure (IOP) at screening: <16 mm Hg, 16 to 23 mm Hg, and >23 mm Hg. IOP was measured with Goldmann applanation tonometry (GAT) as reference method and by ART in both a biomicroscope (ARTBiom) and a handheld (ARTHand) setup with a 10-minutes pause between methods. The mean of 6 readings was regarded as one measurement value.
Results:
Mean age of the subjects was 59 years (range 20 to 87 y). GAT showed a mean IOP of 20.0 mm Hg (range 8.5 to 43.5 mm Hg, n=153). The precision was 2.07 mm Hg for ARTBiom and 2.50 mm Hg for ARTHand, with a significant dependency for age as compared with GAT. Measurement order produced a decreasing IOP with a mean of 2.3 mm Hg between the first and last method.
Conclusions:
The precision obtained in both ARTBiom and ARTHand was within the limits set by the International Standard Organization standards for tonometers. The standardized procedure and the stability of the biomicroscope setup resulted in a slightly better precision as compared with the handheld setup. Despite a 10-minutes pause between measurements, the order was a significant factor, possibly because the patients were more apprehensive at the first measurement.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for: