Related Experiment Video
Updated: Jul 17, 2026

Microvascular Embolism Mouse Model for In Vivo Two-photon Microscopy Using Fluorescent Polystyrene Microspheres
Published on: November 21, 2025
Experimental model evaluation of filter trapping after embolectomy using the Merci system: supplemental technique for
Yasuhiro Suzuki1, Mitsuyuki Fujitsuka, John C Chaloupka
1Department of Radiology, Section of Interventional Neuroradiology, University of Iowa Hospitals and Clinics, Iowa City, IA, USA. CQX05344@nifty.com
Abstract:
Examination of embolectomy using the Merci Retrieval System using experimental stroke models demonstrated that aspiration is not adequate to remove larger clots. The effectiveness of filter trapping was examined using the same models. A silicone model of the carotid artery system with model blood clot was incorporated in a laboratory pulsatile flow system. Embolectomy was performed using the Merci Retrieval System. Any clot not evacuated through the balloon guide catheter was trapped with a distal protection filter device developed for cervical stenting. The clot could not be sucked into the guide catheter by the recommended procedures in nine of 15 trials. Trapping failed in only one trial, in which the clot passed through a gap between the edge of the filter orifice and the inner model lumen. A clot was withdrawn to the catheter tip trapped across the edge of the orifice frame in one trial, and a very large clot was trapped across the filter orifice in two trials. Even clots made by the same method showed variation in properties, especially hardness, which may affect the effectiveness of aspiration. The aspiration procedure recommended for the Merci Retrieval System did not remove the large clots formed by embolectomy. The trapping procedure using a filter device without an orifice frame was effective to solve this problem.
More Related Videos
08:23Implementation of Minimally Invasive Brain Tumor Resection in Rodents for High Viability Tissue Collection
Published on: May 9, 2022
07:05Visualization of Neutrophil Extracellular Traps in Mesenteric Venules After Mesenteric Ischemia-Reperfusion Injury via Intravital Microscopy
Published on: September 27, 2024