X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Impact of additive noise on system performance of a digital X-ray imaging system
Da Zhang1, John Rong, Wei R Chen
1Center for Bioengineering, School of Electrical and Computer Engineering, University of Oklahoma, Norman, OK 73019, USA.
Abstract:
The impact of additive noise on the performance of a digital X-ray imaging system was investigated. The X-ray system is uniquely designed for small animal studies with a focal spot of 20 microm and an adjustable source-to-object distance for radiography. The noise power spectrum and the detective quantum efficiency of this system were measured. The additive noise increased rapidly when the exposure time exceeded a certain range, since the charge-coupled devices of the detector had no cooling system. The noise power spectrum for the additive noise and the noise of the entire imaging system were studied and compared at different exposure times. The detective quantum efficiency was also measured at different exposure times. It was observed that for exposure times less than 10 s, the detective quantum efficiency ((DQE)(0)) is approximately 0.26, dropping to 0.13 at 4 lp/mm and to 0.026 at 8 lp/mm. However, when the exposure exceeds a certain limit (10 s in this study), the rapidly increased additive noise caused the system to be no longer quantum noise limited, resulting in a decreased detective quantum efficiency and a degraded system performance. For example, at an exposure of 20 s, the DQE(O) is approximately 0.22, dropping to 0.11 at 3 lp/mm and to 0.022 at 8 lp/mm.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...