Foreign-body aspiration during microlaryngoscopy: an unusual case of instrument failure
1Department of Otolaryngology, Head and Neck Surgery, The Hospital for Sick Children, University of Toronto, Toronto, Ontario, Canada M5G 1X8.
Abstract:
Instrument failure leading to patient morbidity is a recognized surgical complication that has been reported most commonly in the orthopedic and obstetrical literature [Small NC. Complications in arthroscopy performed by experienced arthroscopists. Arthroscopy 1989;4:215-221; Price MV, Molloy S, Solan MC, Sutton A, Ricketts DM. The rate of instrument breakage during orthopaedic procedures. Int Orthop 2002;26(3):185-7; In Y, Bahk WJ, Park JB. Detachment of the tip of a motorized shaver within the knee joint: a complication of arthroscopic surgery. Arthroscopy 2003;19(6):E29-31; Ostrzenski A. An intraoperative method of localizing a missing piece of a broken laparoscopic instrument. Am J Obstet Gynecol 1997;176:726-7]. Instrument failure resulting in airway foreign bodies, to our knowledge, has been reported only once in the English medical literature [Roach JM, Ripple G, Dillard TA. Inadvertent loss of bronchoscopy instruments in the tracheobronchial tree. Chest 1992;101(2):568-9]. This case report describes the aspiration of a fractured portion of a microlaryngeal instrument caused by metal fatigue and corrosion. The authors recommend vigilant and routine inspection of microlaryngeal instruments for early detection of metal corrosion and fatigue.
Insights
Instrument failure can cause patient harm. This case report highlights a rare instance of a microlaryngeal instrument fracturing due to metal fatigue, leading to airway aspiration, emphasizing the need for regular instrument inspection.
Area of Science:
- Surgical Complications
- Materials Science in Medicine
- Otolaryngology
Background:
- Instrument failure is a known surgical complication, frequently documented in orthopedic and obstetrical procedures.
- Previous reports of instrument failure causing airway foreign bodies are extremely rare, with only one documented case in English literature.
Observation:
- This case report details the aspiration of a fractured microlaryngeal instrument component.
- The instrument failure was attributed to metal fatigue and corrosion.
Findings:
- Microlaryngeal instrument fracture leading to airway aspiration is a critical complication.
- Metal fatigue and corrosion are identified as primary causes of such instrument failure.
Implications:
- Routine and vigilant inspection of microlaryngeal instruments is crucial for early detection of metal corrosion and fatigue.
- Preventing instrument failure can mitigate risks of patient morbidity and rare complications like airway foreign body aspiration.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy Suctioning II: Procedure