Related Experiment Video
Updated: Jul 17, 2026

Home-Based Monitor for Gait and Activity Analysis
Published on: August 8, 2019
Wearable medical devices for tele-home healthcare
1Department of Electronic Engineering, Chinese University of Hong Kong, China.
Abstract:
The world's ageing population and prevalence of chronic diseases have lead to high demand for tele-home healthcare, in which vital-signs monitoring is essential. An overview of state-of-art wearable technologies for remote patient-monitoring is presented, followed by case studies on a cuffless blood pressure meter, ring-type heart rate monitor, and Bluetoothtrade mark-based ECG monitor. Aim of our project is to develop a tele-home healthcare system which utilizes wearable devices, wireless communication technologies, and multisensor data fusion methods. As an important part of this system, a cuffless BP meter has been developed and tested on 30 subjects in a total of 71 trials over a period of five months. Preliminary results show a mean error (ME) of 1.82 mmHg and standard deviation of error (SDE) of 7.62 mmHg in systolic pressure; while ME and SDE in diastolic pressure are 0.45 mmHg and 5.27 mmHg, respectively.
Related Concept Videos
Issues And Trends In Healthcare Delivery System
Cost Containment
Payment for healthcare services has historically promoted adoption of costly and often unnecessary or inefficient...
Documentation in Long-Term and Home Healthcare Setting
Long-Term Care Facilities
Holter Monitor: 24-Hour Monitoring
PPE Use in Healthcare Settings I: Donning
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Integrated Healthcare System