Samar Taha1, Mohamad El-Khatib, Sahar Siddik-Sayyid
1Department of Anesthesiology, American University of Beirut Medical Center, P.O. Box 11 0236 Beirut, Beirut, Lebanon.
Purpose:
This study investigates the efficacy of preoxygenation with Mapleson A and Mapleson D breathing systems vs the circle system with CO2 absorber.
Methods:
Thirteen healthy volunteers underwent tidal volume breathing for three minutes via facemask using Mapleson A, Mapleson D breathing systems or the circle system with CO2 absorber while breathing 100% O2 at flow rates of 5 L.min-1 and 10 L.min-1. Each volunteer acted as his/her own control by going through each of six preoxygenation protocols in random order. Fractional end-tidal O2 concentration (FETO2) was measured at 30-sec intervals. The results were compared among the three anesthesia systems at the two fresh gas flow rates.
Results:
At a fresh gas flow rate of 5 L.min-1, the Mapleson A and circle systems achieved F(ETO2) values of 90.8+/-1.4% and 90.0+/-1.1%, respectively, compared with the lower F(ETO2) (81.5+/-6.3%, P<0.05), achieved with the Mapleson D system. When breathing O2 at 10 L.min-1, the F(ETO2) values after three minutes were similar with the Mapleson A, circle, and Mapleson D breathing systems (91.8+/-2.3%, 91.2+/-1.7%, 90.6+/-2.7%, respectively).
Conclusion:
When using the Mapleson A and the circle systems for preoxygenation, an oxygen flow rate of 5 L.min-1 can adequately preoxygenate the patient within three minutes, while an oxygen flow of 10 L.min-1 is required to achieve a similar fractional end-tidal O2 concentration with the Mapleson D system.
More Related Videos
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
08:26Quantitative Mapping of Specific Ventilation in the Human Lung using Proton Magnetic Resonance Imaging and Oxygen as a Contrast Agent
Published on: June 5, 2019
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...