A multimethod approach to evaluate chemical dot thermometers for oral temperature measurement
Wendy M Fallis1, Kathy Hamelin, Xikui Wang
1University of Manitoba, Winnipeg, Canada. wfallis@vgh.mb.ca
Limited research has explored the accuracy of chemical dot thermometers for oral temperature measurement in adults. This study was undertaken to assess the agreement between oral temperatures taken with an electronic thermometer and single-use chemical dot thermometers in healthy women undergoing surgical delivery. During operative delivery, oral temperatures taken every 15 minutes with both the reference electronic thermometer (ETT) and a Tempa.DOT (TDT) chemical dot thermometer were compared. Data were analyzed using paired t tests, the Bland and Altman plot, and the concordance correlation coefficient. The total number of paired observations for the 62 subjects was 212. The mean difference between the two measurements was 0.35 +/- 0.32 degrees C (p < .0001, 95% CI 0.31, 0.40). Additional analysis indicated a serious undermeasurement by TDT of ETT temperatures. Data from multiple methods of analysis indicate that the Tempa.DOT chemical dot thermometer significantly undermeasures ETT and is not a reliable indicator of oral temperature.
Limited research has explored the accuracy of chemical dot thermometers for oral temperature measurement in adults. This study was undertaken to assess the agreement between oral temperatures taken with an electronic thermometer and single-use chemical dot thermometers in healthy women undergoing surgical delivery. During operative delivery, oral temperatures taken every 15 minutes with both the reference electronic thermometer (ETT) and a Tempa.DOT (TDT) chemical dot thermometer were compared. Data were analyzed using paired t tests, the Bland and Altman plot, and the concordance correlation coefficient. The total number of paired observations for the 62 subjects was 212. The mean difference between the two measurements was 0.35 +/- 0.32 degrees C (p < .0001, 95% CI 0.31, 0.40). Additional analysis indicated a serious undermeasurement by TDT of ETT temperatures. Data from multiple methods of analysis indicate that the Tempa.DOT chemical dot thermometer significantly undermeasures ETT and is not a reliable indicator of oral temperature.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...