Related Experiment Video
Updated: Jul 17, 2026

Demonstration of Equal-Intensity Beam Generation by Dielectric Metasurfaces
Published on: June 7, 2019
Trilateral optical powermeter comparison between NIST, NMIJ/AIST, and METAS
I Vayshenker1, J H Lehman, D J Livigni
1National Institute of Standards and Technology, Colorado 80305, USA. igor@boulder.nist.gov
Abstract:
We describe the results of a comparison of reference standards between three National Metrology Institutes: the National Institute of Standards and Technology (NIST, USA), the National Metrology Institute of Japan/National Institute of Advanced Industrial Science and Technology (NMIJ/AIST, Japan), and the Federal Office of Metrology (METAS, Switzerland). Open-beam- (free field) and optical-fiber-based measurements at wavelengths of 1302 and 1546 nm are reported. Three laboratories' reference standards are compared by means of two temperature-controlled, optical trap detectors. Measurement results show the largest differences of less than 4.2 parts in 10(3), which is within the expanded (k=2) uncertainty for the laboratories' reference standards.