Cuff-less and noninvasive measurements of arterial blood pressure by pulse transit time
1Joint Research Centre for Biomedical Engineering, Department of Electronic Engineering, The Chinese University of Hong Kong, Shatin, Hong Kong.
Insights
A new cuff-less blood pressure (BP) monitoring technique using pulse transit time shows promise. This noninvasive method achieved accuracy comparable to clinical standards, paving the way for wearable BP devices.
Area of Science:
- Biomedical Engineering
- Cardiovascular Technology
- Medical Devices
Background:
- Current blood pressure (BP) measurement devices rely on inflatable cuffs, limiting device miniaturization, power efficiency, and ease of use.
- Cuff-based methods hinder frequent and convenient BP monitoring, impacting self-care and clinical applications.
Purpose of the Study:
- To develop and evaluate a novel, cuff-less, noninvasive technique for measuring blood pressure (BP) utilizing pulse transit time (PTT).
- To assess the feasibility of PTT-based BP measurement for potential integration into wearable health monitoring devices.
Main Methods:
- The study involved 85 participants (36 males, 39 hypertensive, average age 57±29 years) over 6.4 weeks.
- A total of 999 pairs of systolic BP (SBP) and diastolic BP (DBP) estimations were performed using the PTT technique.
- Reference BP values were obtained from an experienced registered nurse and a clinically approved automatic BP meter.
Main Results:
- The cuff-less PTT technique demonstrated a mean difference of 0.6±9.8 mmHg for SBP and 0.9±5.6 mmHg for DBP compared to reference measurements.
- These results approach the Association for the Advancement of Medical Instrumentation (AAMI) requirements of ±5±8 mmHg for BP measurement accuracy.
Conclusions:
- The developed cuff-less BP measurement technology exhibits significant potential for creating advanced wearable devices.
- This innovation could greatly benefit home healthcare through self-monitoring and eventually aid in clinical diagnosis.
- The PTT-based approach offers a viable alternative to traditional cuff-based BP monitors, enhancing user convenience and device applicability.
Abstract:
The current blood pressure (BP) measurement devices are mostly built on the principle of auscultation, oscillometry or tonometry, all of which use an inflatable cuff to occlude or unload the artery. The need of a cuff in these devices limits the further reduction in size and power consumption, and restricts the frequency and ease of their usage. Therefore, this study aims to develop a cuff-less and noninvasive technique for measuring BP by pulse transit time. The technique was evaluated on 85 subjects, aged 57+/-29 yrs., including 36 males and 39 hypertensives, over an average period of 6.4 wks. A total of 999 pairs of systolic BP (SBP) and diastolic BP (DBP) estimations were made. The average of BP readings reported by an experienced registered nurse and those obtained from a clinically approved automatic BP meter was used as reference. It is found that the estimated SBP and DBP differed from the reference BP by 0.6+/-9.8mmHg and 0.9+/-5.6mmHg respectively. When compared to the AAMI requirement (5+/-8mmHg for both SBP and DBP estimation), the results suggest that the cuff-less technology has great potential to be developed into wearable devices that are useful in self BP monitoring for home healthcare and eventually in clinical diagnosis.
More Related Videos
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical indicator...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessing Blood pressure in the Leg
Preparation: