High precision Multifrequency Electrical Impedance Tomography System and Preliminary imaging results on saline tank
Shi Xuetao1, You Fusheng, Fu Feng
1Medical Electronic Engineering Department, Fourth Military Medical University, Xi'an, 710033, China.
Abstract:
To establish a high precision data acquisition system for multi-frequency electrical impedance tomography (EIT), a series of methods were introduced. Those methods include building a driving signal with up to four frequency components to diminish the effect of the dynamic change of tissues resistivity, extracting the impedance information by a digital demodulator that can improve the SNR by 8 times. The system that established can work at a wide range from 1.6kHz to 380kHz. Its CMRR is 74dB at 100kHz. The output impedance of current source is 2MΩ at that frequency. And measurement precision on a 100ohm resistor is better than -80dB in full bandwidth. Both the quasi-static and the dynamic imaging results based on a saline tank can reflect the resistivity changes inside the phantom clearly. Therefore, the system was competent in multifrequency EIT research work.