The sensor in the venous chamber does not prevent passage of air bubbles during hemodialysis
Bernd Stegmayr1, Ulf Forsberg, Per Jonsson
1Division of Nephrology, Department of Internal Medicine, University of Umeå, Umeå, Sweden. bernd.stegmayr@medicin.umu.se
Abstract:
We previously showed, in vitro, that micro bubbles pass the air trap without inducing an alarm. The aim was to investigate if micro bubbles bypass the detector during hemodialysis (HD). During HD (40 patients, 47 HD sessions, 231 measurements), an ultrasound detector was fixed just after the venous air trap. Micro bubble size was measured in the range from 5 microm up to >42.5 microm. Blood flow was at a mean 346 mL/min (SD +/- 57). The mean of all micro bubbles per minute, without inducing an alarm, was at start 128 (range 0-769). Measurements revealed the presence of micro bubbles in all of the series and in 90% of the measurements. There was no difference between start and end of the same dialyses. There was a correlation between blood flow and extent of micro bubbles for the smaller sizes and the sum of all bubbles (r > or = 0.29, P < or = 0.026). Micro bubbles passed the air trap without alarming. Most bubbles were approximately 5 microm.
Related Concept Videos
Hemodialysis I: Introduction
Hemodialysis II: Procedure and Complications
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis III: Nursing Management
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy