Double-wire technique in balloon-protected carotid artery stenting
Osami Kawarada1, Yoshiaki Yokoi, Kazushi Takemoto
1Department of Cardiology, Kishiwada Tokushukai Hospital, Osaka, Japan. kawarada90@hotmail.com
Journal of Interventional Cardiology
|February 16, 2007
Summary
The double-wire technique for carotid artery stenting (CAS) is safe and effective, reducing occlusion time. This method, using the PercuSurge GuardWire, offers satisfactory outcomes for patients.
Area of Science:
- Vascular Surgery
- Interventional Cardiology
- Medical Device Technology
Background:
- Carotid artery stenting (CAS) is a crucial procedure for stroke prevention.
- Distal occlusion balloon systems aim to minimize embolic events during CAS.
- Optimizing procedural efficiency and safety remains a key focus in CAS.
Purpose of the Study:
- To evaluate the safety and efficacy of a novel "double-wire technique" for balloon-protected CAS.
- To assess the performance of the PercuSurge GuardWire system with this technique.
- To determine the impact of the double-wire technique on cerebral blood flow occlusion time.
Main Methods:
- A prospective study involving 51 patients with 53 de novo carotid artery lesions.
- Implementation of the "double-wire technique" utilizing one wire for predilatation and another for stent implantation.
- Use of the PercuSurge GuardWire distal occlusion balloon system during the procedure.
Main Results:
- Technical success was achieved in all 50 treated lesions.
- Cerebral blood flow occlusion time averaged 266 +/- 37 seconds, with 90% of procedures under 300 seconds.
- Low rates of neurological intolerance (7.7%) and periprocedural complications (3.8%) were observed.
Conclusions:
- The double-wire technique is a safe and effective approach for balloon-protected CAS.
- This technique successfully shortens occlusion time, leading to favorable technical and clinical results.
- The double-wire technique using the PercuSurge GuardWire is recommended for consideration in CAS procedures.