A Point-of-Care Method with Integrated Decision Support Tool to Estimate Anemia at Population Level
Published on: January 19, 2024
Point of care estimation of haemoglobin in neonates
Lucy E Hinds1, Caroline L Brown, Simon J Clark
1Paediatric Department, Barnsley Hospital Foundation Trust, Barnsley, UK. markandlucyhinds@gmail.com
Objective:
To evaluate whether measurement of haemoglobin concentration in neonates using point of care testing agrees with laboratory measurement.
Design:
127 paired blood samples taken from babies on a neonatal intensive care unit for full blood count and blood gas analysis by point of care testing were reviewed according to current practice. A comparison was made between the laboratory and blood gas analyser haemoglobin measurements to assess limits of agreement and look for any systematic difference.
Setting:
Neonatal Unit, Jessop Wing, Royal Hallamshire Hospital, Sheffield, UK PATIENTS: Babies staying on the neonatal unit, who currently have contemporaneous blood samples taken for full blood count and blood gas analysis by point of care testing.
Intervention:
Results from blood samples were reviewed.
Main Outcome Measure:
Comparison between laboratory and point of care testing haemoglobin concentrations.
Results:
The mean laboratory haemoglobin concentration was 155 g/l (range 30-226 g/l); the mean point of care testing haemoglobin concentration was 157 g/l (range 30-228 g/l). The mean (SD) difference between paired samples was 2 (11) g/l; 95% CI -4.0 to 0.1 g/l; and limits of agreement -23 to 19 g/l.
Conclusions:
The blood gas analyser on the neonatal unit at Royal Hallamshire Hospital, Sheffield, gives a useful estimation of haemoglobin concentration compared with laboratory measurement, with smaller sample volume. Although this does not replace a full blood count, it is a useful adjunct to neonatal care monitoring.