Related Experiment Video
Updated: Jul 16, 2026

A Real-World High-Intensity Interval Training Protocol for Cardiorespiratory Fitness Improvement
Published on: February 22, 2022
Monitoring VO2max during fourteen weeks of endurance training using the CardioCoach
Pat R Vehrs1, David M Keller, James D George
1Department of Physical Education, Brigham Young University, Provo, Utah 84602, USA. pat_vehrs@byu.edu
Abstract:
This study evaluated the validity of the desktop CardioCoach metabolic system to measure VO2max and VEmax. Sixteen subjects (mean age = 19.5 +/- 3.2 years) completed 2 maximal graded exercise tests following the same protocol before and after 7 and 14 weeks of endurance training. Subjects' VO2max and VEmax were measured by either the CardioCoach or the ParvoMedics TrueOne 2400 metabolic measurement system (TrueOne). An alpha level of significance of p < 0.05 was maintained for all statistical analyses. The time to test completion and the final treadmill grade of the exercise tests performed by both the CardioCoach and the TrueOne increased over the 3 testing periods, confirming an improvement in cardiorespiratory fitness resulting from the 14 weeks of training. A linear growth curve analysis indicated that there were statistically significant differences between VO2max (ml x kg(-1) x min(-1)) as measured by the TrueOne and the CardioCoach before (44.4 +/- 5.0 and 49.3 +/- 5.4) and after 7 weeks (46.0 +/- 5.2 and 48.2 +/- 5.4) of training but not after 14 weeks of training (47.8 +/- 5.6 and 48.4 +/- 5.2). Significant differences also existed in VEmax (L x min(-1)) as measured by the TrueOne and the CardioCoach before (76.8 +/- 17.7 and 71.9 +/- 13.7), after 7 weeks (81.4 +/- 16.2 and 72.8 +/- 14.1), and after 14 weeks (86.8 +/- 19.4 and 74.2 +/- 13.1) of training. Although significant growth of VO2max (0.24 ml x kg(-1) x min(-1) x wk(-1)) and VEmax (0.71 L x min(-1) x wk(-1)) was measured by the TrueOne over 14 weeks of training, the CardioCoach was unable to detect growth in VO2max (-0.02 ml x kg(-1) x min(-1) x wk(-1)) or VEmax (0.17 L x min(-1) x wk(-1)). This study indicates that the CardioCoach did not accurately measure or monitor changes in VO2max or VEmax resulting from training.
More Related Videos
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
Related Concept Videos
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...
Holter Monitor: 24-Hour Monitoring
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Exercise and Cardiovascular Response
Light to moderate physical activity initiates a series of interconnected responses in the body. The heart rate modestly increases in anticipation of the workout, followed by widespread vasodilation as oxygen consumption by skeletal muscles increases. This results in decreased peripheral resistance, increased capillary blood flow, and accelerated...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...