Related Experiment Video
Updated: Jul 16, 2026

A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
Non-invasive assessment of endothelial function. Intra and inter-observer variability
Arturo Sotomayor González1, Andrea Kostine, Jorge R Gómez-Flores
1Electrocardiology Department, Instituto Nacional de Cardiologia lgnacio Chávez, México, D.F
Background And Objectives:
Non-invasive evaluation of endothelial function with high resolution ultrasound has become a widely accepted tool in determination of high risk subjects for early atherosclerosis. Despite its simple appearance, ultrasonographic assessment of brachial artery changes, is technically challenging and has a significant learning curve. In the present study, we evaluate the intra and inter-observer variability in assessing peripheral endothelial function with high resolution ultrasound at a tertiary referral center.
Methods:
Assessment of endothelial function was performed by 2 physicians in 20 volunteers without evidence of coronary artery disease. Endothelial function is evaluated with a high frequency bidimensional ultrasound with a 10.0-MHz linear-array transducer used for the study. Each volunteer was examined by both observers using an identical protocol, measuring brachial artery diameter on three occasions.
Results:
Excellent correlation was observed for all brachial artery measures with a Spearman's correlation coefficient > 0.9 (p < 0.0001). Flow-mediated dilation (FMD) in the study population was of 10.45+/-6.9%. Flow independent dilation (FID) was of 24.35+/-7.63%. Intra-observer variability was of 2.1% for observer A and 1.1 % for observer B.
Conclusions:
Non-invasive assessment of endothelial function using brachial artery ultrasound is reproducible and can be performed with low intra and inter-observer variability.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: