Percutaneous transtracheal ventilation: resuscitation bags do not provide adequate ventilation
Edmond A Hooker1, Daniel F Danzl, Daniel O'Brien
1Department of Health Services Administration, Xavier University, Cincinnati, OH 45207-7331, USA. hookere@xavier.edu
Introduction:
Percutaneous, transtracheal jet ventilation (PTJV) is an effective way to ventilate both adults and children. However, some authors suggest that a resuscitation bag can be utilized to ventilate through a cannula placed into the trachea.
Hypothesis:
Percutaneous transtracheal ventilation (PTV) through a 14-gauge catheter is ineffective when attempted using a resuscitation bag.
Methods:
Eight insufflation methods were studied. A 14-gauge intravenous catheter was attached to an adult resuscitation bag, a pediatric resuscitation bag, wall-source (wall) oxygen, portable-tank oxygen with a regulator, and a jet ventilator (JV) at two flow rates. The resuscitation bags were connected to the 14-gauge catheter using a 7 mm adult endotracheal tube adaptor connected to a 3 cc syringe barrel. The wall and tank oxygen were connected to the 14-gauge catheter using a three-way stopcock. The wall oxygen was tested with the regulator set at 15 liters per minute (LPM) and with the regulator wide open. The tank was tested with the regulator set at 15 and 25 LPM. The JV was connected directly to the 14-gauge catheter using JV tubing supplied by the manufacturer. Flow was measured using an Ohmeda 5420 Volume Monitor. A total of 30 measurements were taken, each during four seconds of insufflation, and the results averaged (milliliters (ml) per second (sec)) for each device.
Results:
Flow rates obtained using both resuscitation bags, tank oxygen, and regulated wall oxygen were extremely low (adult 215 +/- 20 ml/sec; pediatric 195 +/- 19 ml/sec; tank 358 +/- 13 ml/sec; wall at 15 l/min 346 +/- 20 ml/sec). Flow rates of 1,394 +/- 13 ml were obtained using wall oxygen with the regulator wide open. Using the JV with the regulator set at 50 pounds per square inch (psi), a flow rate of 1,759 +/- 40 was obtained. These were the only two methods that produced flow rates high enough to provide an adequate tidal volume to an adult.
Conclusions:
Resuscitation bags should not be used to ventilate adult patients through a 14-gauge, transtracheal catheter. Jet ventilation is needed when percutaneous transtracheal ventilation is attempted. If jet ventilation is attempted using oxygen supply tubing, it must be connected to an unregulated oxygen source of at least 50 psi.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...