Related Experiment Video
Updated: Jul 16, 2026

Biomolecular Detection employing the Interferometric Reflectance Imaging Sensor (IRIS)
Published on: May 3, 2011
Oblique-incidence reflectivity difference microscope for label-free high-throughput detection of biochemical
Xiangdong Zhu1, James P Landry, Yung-Shin Sun
1Department of Physics, University of California, CA 95616, USA. xdzhu@physics.ucdavis.edu
Abstract:
We describe a recently developed oblique-incidence reflectivity difference (OI-RD) microscope, a form of polarization-modulated imaging ellipsometer, for label-free-high-throughput detection of biomolecular reactions on DNA and protein microarrays. We present examples of application of this technique to end-point and real-time investigations of DNA-DNA hybridization, antibody-antigen capture, and protein-small-molecule binding reactions. Compared to a conventional imaging ellipsometer based on the polarizer-compensator-sample-analyzer scheme and under the off-null condition, a polarization-modulated OI-RD microscope is inherently more sensitive by at least 1 order of magnitude to thickness changes on a solid surface. Compared with imaging surface plasmon resonance microscopes based on reflectance change on falling or rising slopes of the surface plasmon resonance, the OI-RD microscope (1) has a comparable sensitivity, (2) is applicable to conventional microscope glass slides, and (3) easily covers a field of view as large as the entire surface of a 1 in. x 3 in. (2.54 cm x 7.62 cm) microscope slide.

