Related Experiment Video
Updated: Jul 16, 2026

Design of a Biocompatible Drug-Eluting Tracheal Stent in Mice with Laryngotracheal Stenosis
Published on: January 21, 2020
Benign tracheobronchial strictures: long-term results and factors affecting airway patency after temporary stent
Jin Hyoung Kim1, Ji Hoon Shin, Ho-Young Song
1Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, 388-1, Poongnap-dong, Songpa-gu, Seoul 138-736, South Korea.
Objective:
The purpose of our study was to evaluate long-term results and identify factors affecting airway patency after temporary placement of a covered, retrievable nitinol stent for benign tracheobronchial strictures.
Materials And Methods:
Polyurethane or polytetrafluoroethylene (PTFE)-covered retrievable expandable nitinol stents were placed fluoroscopically in 24 patients with benign tracheobronchial strictures. Improvement in respiratory status and complications were evaluated. Maintained patency of airway after temporary stenting was calculated and compared between the 2- and 6-month stenting groups. Factors for maintained patency after temporary stenting were evaluated.
Results:
A total of 30 stents were successfully placed and well tolerated in 24 patients. Tissue hyperplasia, stent migration, and bronchial obstruction of the left upper lobe occurred in 36.7%, 13.3%, and 3.3% of patients, respectively. All stents were successfully removed electively either 2 (n = 12) or 6 (n = 12) months after placement or when complications occurred (n = 6). During the follow-up period (mean, 24 months), dyspnea recurred in 15 of the 24 patients. The 6-month stenting group showed a lower recurrence rate (41.7% vs 83.3%, p = 0.045) and a better mean maintained patency (39.7 +/- 7.8 vs 9.4 +/- 5.4 months, p = 0.001) than the 2-month stenting group. Multivariate analysis showed that duration of stent placement (p = 0.002) and the occurrence of tissue hyperplasia (p = 0.026) were associated with maintained patency after temporary stenting.
Conclusion:
Temporary placement of a covered, retrievable, expandable nitinol stent may be a safe and effective treatment for benign tracheobronchial strictures during the period the stent is in place. A high symptomatic recurrence rate of 62.5% was found after stent removal. Shortterm placement of the stent and tissue hyperplasia were associated with decreased airway patency.
Related Concept Videos
Esophageal Strictures-I: Introduction
Etiology
The primary cause of esophageal strictures is long-standing gastroesophageal reflux disease (GERD), accounting for about 70 to 80% of adult cases. Chronic acid reflux can lead to injury and scarring of the esophageal lining, culminating in...
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.