Related Experiment Video
Updated: Jul 16, 2026

Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Homogeneous enhancement in pediatric thoracic CT aortography using a novel and reproducible method: contrast-covering
I-Chen Tsai1, Tain Lee, Min-Chi Chen
1Department of Radiology, Taichung Veterans General Hospital, No. 160, Sec. 3, Taichung Harbor Rd., Taichung 407, Taiwan, ROC.
Objective:
This study compares the empiric setting and contrast-covering time (CCT) concept for i.v. contrast injection in pediatric thoracic CT aortography.
Subjects And Methods:
A total of 113 pediatric patients referred for thoracic CT aortography were classified into four groups: group 1 (1-5 years old, CCT), group 2A (6-10 years old, CCT), group 2B (6-10 years old, empiric setting), and group 3 (11-15 years old, CCT). The CT attenuation values from the left common carotid artery to the descending aorta were recorded every 0.5 second. The quantitative bolus geometric analysis of each group included average enhancement, SD within the patient, and slope of enhancement. Groups 2A and 2B were compared to determine whether better bolus geometry could be obtained with the CCT concept than with the traditional empiric setting. Groups 1, 2A, and 3 were compared to determine whether homogeneous bolus geometry could be obtained in different age groups.
Results:
More homogeneous enhancement was obtained with the CCT concept than the empiric setting with a smaller SD of enhancement (25.5 +/- 8.5 H vs 49.3 +/- 16.2 H, p < 0.001). Furthermore, in the three different age groups (groups 1, 2A, and 3) examined using the CCT concept, there was no significant difference in the average enhancement (415.7 +/- 83.6 H, 422.8 +/- 97.1 H, 392.0 +/- 78.5 H, respectively; all p > 0.05), SD of enhancement (28.5 +/- 9.8 H, 25.5 +/- 8.5 H, 28.5 +/- 14.6 H, respectively; all p > 0.05), or enhancement slopes (-5.6 +/- 18.0 H, -2.7 +/- 10.7 H, -5.4 +/- 12.3 H, respectively; all p > 0.05).
Conclusion:
The CCT concept yields more homogeneous enhancement than the empiric setting. It also can routinely obtain homogeneous bolus geometry in patients in different age groups.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...