Related Experiment Video
Updated: Jul 16, 2026

An “All-laser” Endothelial Transplant
Published on: July 6, 2015
"Top hat"-shaped penetrating keratoplasty using the femtosecond laser
Roger F Steinert1, Teresa S Ignacio, Melvin A Sarayba
1Department of Ophthalmology, University of California Irvine, California 92697, USA. steinert@uci.edu <steinert@uci.edu>
Purpose:
To determine the feasibility of using a modified femtosecond laser to perform penetrating corneal resections.
Design:
Controlled laboratory study.
Methods:
Fourteen corneoscleral rims were mounted on an artificial anterior chamber. Eight underwent traditional penetrating keratoplasty (PKP) and six underwent "top hat"-shaped PKP using a femtosecond laser. Wound leakage pressure and induced astigmatism were compared between the two groups.
Results:
Deep corneal resection with the femtosecond laser was feasible. Wound leakage occurred at 38 +/- 11 mm Hg in the traditional PKP eyes and at 240 +/- 69 mm Hg in the laser-shaped PKP groups. Initial induced astigmatism was 3.76 +/- 0.82 diopters and 3.46 +/- 1.36 diopters in the traditional and shaped PKP groups, respectively.
Conclusions:
Shaped PKP using the femtosecond laser is feasible and provides superior incision integrity compared to traditional PKP. Initial induced astigmatism is dominated by suture effects.