Technical note: overprediction of dose with default PRESTA-I boundary crossing in DOSXYZnrc and BEAMnrc
1Ionizing Radiation Standards, National Research Council of Canada, Ottawa, KIA OR6 Canada. bwalters@irs.phy.nrc.ca
Abstract:
Use of the PRESTA-I (current default) boundary crossing algorithm for charged particles in DOSXYZnrc and BEAMnrc is shown to result in dose overprediction by up to 2.5% in photon and electron beams (less in high-energy electron beams). The effect is most pronounced when charged particle equilibrium does not exist or when the phantom is not divided into uniform voxels throughout. As a result, the default boundary crossing algorithm in BEAMnrc will be changed to "exact" (single scattering).
More Related Videos
Related Concept Videos
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force per...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Dosage Compensation
In addition to sexual development, the X chromosome has genes involved in autosomal functions such as brain development and the immune system. Therefore, males and females with distinct numbers of X chromosomes will have...
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady state,...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations