Sources of long-term variability in measurements of lung function: implications for interpretation and clinical trial
Robert L Jensen1, John G Teeter, Richard D England
1Pulmonary Laboratory, LDS Hospital and University of Utah, Eighth Ave and C St, Salt Lake City, UT 84143, USA. ldrjens1@ihc.com
Background:
The objective of the study was to characterize the biological and technical components of variability associated with longitudinal measurements of FEV(1) and carbon monoxide diffusing capacity (Dlco). Variability was apportioned to subject and instrument for five commercially available pulmonary function testing (PFT) systems: Collins CPL (Ferraris Respiratory; Louisville, CO); Morgan Transflow Test PFT System (Morgan Scientific; Haverhill, MA); SensorMedics Vmax 22D (VIASYS Healthcare; Yorba Linda, CA); Jaeger USA Masterscreen Diffusion TP (VIASYS Healthcare; Yorba Linda, CA); and Medical Graphics Profiler DX System (Medical Graphics Corporation; St. Paul, MN).
Methods:
This was a randomized, replicated cross-over, single-center methodology study in 11 healthy subjects aged 20 to 65 years. Spirometry and Dlco measurements were performed at baseline, 3 months, and 6 months. Repetitive simulations of FEV(1) and Dlco were performed on the same instruments on four occasions over a 90-day period using a spirometry waveform generator and a Dlco simulator.
Results:
The coefficient of variation associated with repetitive measurements of FEV(1) or Dlco in subjects was consistently larger than that associated with repetitive simulated waveforms across the five instruments. Instrumentation accounted for 13 to 58% of the total FEV(1) and 36 to 70% of the total Dlco variability observed in subjects. Sample size estimates of hypothetical studies designed to detect treatment group differences of 0.050 L in FEV(1) and 0.5 mL/min/mm Hg in Dlco varied as much as four times depending on the instrument utilized.
Conclusions:
These results provide a semiquantitative assessment of the biological and technical components of PFT variability in a highly standardized setting. They illustrate how instrument choice and test variability can impact sample size determinations in clinical studies that use FEV(1) and Dlco as end points.
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Lung Capacity
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...