Measurement of Elecsys NT-proBNP in serum, K2 EDTA and heparin plasma
Giuseppe Lippi1, Gian Luca Salvagno, Martina Montagnana
1Sezione di Chimica e Microscopia Clinica, Dipartimento di Scienze Morfologico-Biomediche, Università degli Studi di Verona, Ospedale Policlinico G.B. Rossi, Piazzale Scuro, 10, 37134-Verona, Italy. giuseppe.lippi@univr.it
Objective:
There is controversial evidence on a matrix influence on the measurement of NT-proBNP.
Design And Methods:
We compared results of Elecsys NT-proBNP measurement on serum, K2 EDTA plasma and lithium heparin plasma.
Results:
Samples collected in K2 EDTA showed a marginally significant underestimation when compared to serum and heparin, whereas no significant difference was observed between serum and heparin plasma.
Conclusions:
Serum, heparin and K2 EDTA plasma may be suitable for NT-proBNP measurement.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...