High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
Published on: June 27, 2025
Minimal sensor count approach to fuzzy logic rotary blood pump flow control
Fernando Casas1, Nisar Ahmed, Andrew Reeves
1Lerner Research Institute, Department of Biomedical Engineering, The Cleveland Clinic Foundation, Cleveland, OH 44195, USA.
Abstract:
A rotary blood pump fuzzy logic flow controller without flow sensors was developed and tested in vitro. The controller, implemented in LabView, was set to maintain a flow set point in the presence of external pressure disturbances. Flow was estimated as a function of measured pump's delta P and speed, using a steady-state, nonlinear approximation. The fuzzy controller used the pump's flow estimate and delta P as feedback variables. The defuzzified control output manipulated the pump speed. Membership functions included flow error, delta P, and pump speed. Experimental runs in a mock loop (water/glycerin 3.5 cPs, 37 degrees C), using the estimated flow, were compared with those using a Transonic flow meter for nine conditions of flow and delta P (4 to 6 L/min, 150 to 350 mm Hg). Pressure disturbances generated by a servo pinch valve ranged from +/-23 to +/-47 mm Hg. Results indicated that the fuzzy controller ably regulated the flow set point to within +/-10% of the baseline even under large swings in pressure. There was no difference in controller performance between the ultrasonic flow measurement and the estimated flow calculation scenarios. These tests demonstrated that the fuzzy controller is capable of rejecting disturbances and regulating flow to acceptable limits while using a flow estimate.
More Related Videos
07:34A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
09:08Three-dimensional Printing of Thermoplastic Materials to Create Automated Syringe Pumps with Feedback Control for Microfluidic Applications
Published on: August 30, 2018
Related Concept Videos
Pipe Flowrate Measurement: Problem Solving
Pipe Flowrate Measurement
The orifice meter is a simple,...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
PID Controller
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal and...
Bioreactor Controls-I