Is pulse synchronized pneumotonometry more reproducible than routine pneumotonometry and more in agreement with
A-S Lafaut1, L Van Malderen, T Zeyen
1Department of Ophthalmology, University Hospitals Leuven, 3000 Leuven, Belgium.
Purpose:
This study compared the variability of intraocular pressure (IOP) measurements taken with pulse synchronized pneumotonometry (PNT) and with routine PNT (without pulse synchronization) and evaluated the agreement between PNT and Goldmann applanation tonometry (GAT).
Methods:
In this prospective study, 148 eyes from 78 patients were enrolled. Patients were randomized into two groups. In the first group (A), the sequence of measurements was pulse synchronized PNT, routine PNT, and GAT. In the second group (B), the sequence of measurements was routine PNT, pulse synchronized PNT, and GAT. The mean of three measurements was averaged for PNT and GAT. All the measurements were performed by the same investigator, who was masked to GAT measurements. The mean IOP measurements and intrapatient standard deviations among the three tonometers were compared using analysis of variance measurement. Bland & Altman plots were used to assess the agreement between PNT and GAT.
Results:
The variability of IOP measurements taken with pulse synchronized PNT was significantly lower than with routine PNT in both groups (1.2+/-0.7 vs 1.3+/-0.7 mmHg in group A and 1.1+/-0.9 vs 1.3+/-1.0 mmHg in group B [p<0.001], respectively). PNT measurements with and without pulse synchronization were on average +/-2 mmHg higher than GAT measurements in both groups (p<0.001). The 95% limits of agreement between PNT and GAT varied between -3.8 and 8.5 mmHg.
Conclusions:
Pulse synchronized PNT gives more reproducible measurements than routine PNT. The agreement between PNT and GAT is poor.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Measurement of Blood Pressure
Special considerations while measuring pulse
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.