Related Experiment Video
Updated: Jul 15, 2026

Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Non-invasive ultrasonic cardiac output monitoring in exercise testing
Background:
The purpose of this study was to evaluate the relationship of breath-by-breath oxygen uptake (VO(2)) and CO, cardiac index (CI), stroke volume (SV) and heart rate (HR) determined by a non-invasive continuous wave (CW) Doppler-based system (USCOM) in healthy adolescent and adult athletes during exercise treadmill testing.
Methods:
Twenty athletes were enrolled for a standardized ramp treadmill test including transthoracic echocardiography at rest, incremental breath-by-breath spiroergometry, lactate kinetics and CW Doppler.
Results:
VO(2) max was 51+/-11 ml/min/kg in the adolescent cross country skiers (CI 8.2+/-1.2 l/min/m(2)) and 50+/-12 ml/min/kg in adult soccer referees (CI 7.1+/-1.2 l/min/m(2)). CW Doppler hemodynamic variables showed significant correlation with VO(2) (p<0.01, n=95) across the total of measures, with excellent correlation demonstrated for CO and CI (0.904 and 0.897), while HR and SV showed good to very good correlation with r values of 0.842 and 0.766, respectively. Correlation of the variables at peak exercise and 1 and 3 min after recovery, demonstrate that CO and SV show superior correlation with peak oxygen consumption (VO(2) max). The regression equation for VO(2) and CO(USCOM) was VO(2)=4.43CO(USCOM)-10.3.
Conclusion:
CW Doppler measurement of cardiac output is feasible as a measure of cardiac performance during standardized treadmill testing in adolescent and adult athletes.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Exercise and Cardiac Output
Sustained exercise increases the muscles' oxygen demand, which can be met...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes