Systematic errors and susceptibility to noise of four methods for calculating anatomical dead space from the CO2
Y Tang1, M J Turner, A B Baker
1Department of Anaesthetics, Royal Prince Alfred Hospital, University of Sydney, NSW 2050, Sydney, Australia. ytan7763@usyd.edu.au
Background:
Anatomical dead space is usually measured using the Fowler equal area method. Alternative methods include the Hatch, Cumming, and Bowes methods, in which first, second, and third order polynomials, respectively, fitted to an expired CO2 volume vs expired volume curve, intercept the x-axis at the anatomical dead space. This study assessed systematic errors and susceptibility to noise of the Fowler, Hatch, Cumming, and Bowes dead spaces calculated over 40-80% of the CO2 expirogram.
Methods:
Simulated CO2 expirograms with 220 ml anatomical dead space and varying alveolar plateau slopes were generated digitally and zero-mean Gaussian noise added. CO2 expirograms were recorded in 10 anaesthetized human subjects. Anatomical dead space was calculated by the Fowler, Hatch, Cumming, and Bowes methods.
Results:
The Fowler, Hatch, Cumming, and Bowes methods displayed systematic biases of -1.8%, 13.2%, 2.4%, and -1.3%, respectively, at a normalized simulated alveolar plateau slope of 1.6 litre(-1). At a noise level of 0.0066 vol/vol, the standard deviations of recovered simulated dead spaces were 70.6, 1.8, 2.4, and 3.7 ml, respectively. The Hatch, Cumming, and Bowes methods applied to human expirograms differed significantly from that of Fowler by 13, -4, and -11 ml, respectively. In the human study, the Hatch and Cumming methods yielded the lowest intra-individual dead space variability.
Conclusions:
The Fowler method shows greatest susceptibility to measurement noise and the Hatch method exhibits the largest systematic error. The Cumming method, which exhibits both low bias and low noise susceptibility, is preferred for estimating anatomical dead space from CO2 expirograms.
More Related Videos
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Random and Systematic Errors
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Random and Systematic Errors