Simple technique for obtaining photoacoustic spectra corrected for the spectral variation of the source in single
K Sathiyamoorthy1, C Vijayan, M P Kothiyal
1Department of Physics, Indian Institute of Technology Madras, Chennai, India.
Abstract:
A simple technique is presented to obtain normalized photoacoustic (PA) spectra corrected for the spectral variation of the source in a single scan. The input light beam is passed through the center of a dual slot chopper, which splits it into two chopped output beams at two different frequencies at a fixed ratio. The beams fall on the sample and the reference kept side-by-side in the same sample chamber. The PA signals are detected by a microphone and processed by two lock-in amplifiers tuned at two different frequencies. The technique is tested by recording the PA spectra of standard samples.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Interference
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...