Related Experiment Video
Updated: Jul 15, 2026

Synthesis and Purification of Iodoaziridines Involving Quantitative Selection of the Optimal Stationary Phase for Chromatography
Published on: May 16, 2014
Flow injection determination of anisidine value in palm oil samples using a triiodide potentiometric detector
Bahruddin Saad1, Wan Tatt Wai, Boey Peng Lim
1School of Chemical Sciences, Universiti Sains Malaysia, 11800 Penang, Malaysia. bahrud@usm.my
Abstract:
A flow injection analysis (FIA) procedure for the determination of anisidine value (AV) in palm olein using a triiodide detector is described. Undiluted oil sample and chloramine-T reagent were added to a reaction chamber, and reaction was accelerated by applying a short vortex action (typically for 30 s). After allowing the emulsified oil phase to be separated from the aqueous phase (bottom layer), an aliquot of the aqueous phase (containing unreacted chloramine-T) was aspirated into a carrier stream that contained I(-) where the chloramine-T oxidized the I- to form I3(-) which was finally detected by a flow-through triiodide potentiometric detector. Variables that affect the FIA signals such as size of the reaction chamber, oil and reagent flow rates, chloramine-T concentration, vortex time, time for phase separation, carrier stream pH and injected volume were studied. The optimized FIA procedure is linear over 1.0-23.0 AV. The method exhibits good repeatabililty (R.S.D. of +/-3.16% (n = 4) for the determination of 5.0 AV) and a sampling rate of 40 samples per hour was achieved. Good correlation (r2 = 0.996 (n = 4)) between the proposed method and the manual American Oil Chemists' Society procedure was found when applied to the determination of twenty different types of palm olein samples.
Related Concept Videos
Redox Titration: Iodimetry and Iodometry
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Redox Titration: Overview