Morphological and Functional Assessment of the Right Ventricle Using 3D Echocardiography
Published on: October 28, 2020
Variables influencing the accuracy of right ventricular volume assessment by real-time 3-dimensional
Martin Hoch1, Nikolay V Vasilyev, Brian Soriano
1Department of Cardiology, Children's Hospital Boston, Boston, Massachusetts 02115, USA.
Background:
Preliminary experience with matrix-array real-time (RT) 3-dimensional echocardiography (3DE) in pediatric patients has demonstrated consistently lower right ventricular volume (RVV) compared with magnetic resonance imaging. Our hypothesis was that variables in RT 3DE acquisition and offline analysis, including gain settings, thickness, and orientation of disks, could alter RVV measurements.
Methods:
Displacements of water from latex models derived from excised lamb hearts were used for comparison. RT 3DE volume acquisitions were performed using a matrix-array probe (2-4 MHz); RVVs were calculated offline using summation of disks method.
Results:
No significant difference and excellent agreement was found for comparison of RT 3DE with displacement of water using 5-mm cut planes, optimal gain settings, and short-axis tracings. Different gain settings and long-axis tracings significantly affected RVV. A slice thickness of 13 mm or greater affected volume measurements.
Conclusions:
RT 3DE can accurately measure RVV. Specific variables will alter volumes measurements and must be considered in clinical studies.