Related Experiment Video
Updated: Jul 15, 2026

Optimized Griess Reaction for UV-Vis and Naked-eye Determination of Anti-malarial Primaquine
Published on: October 11, 2019
Rapid and sensitive HPLC method for the simultaneous determination of paraquat and diquat in human serum
Shuuji Hara1, Noriaki Sasaki, Daizi Takase
1Faculty of Pharmaceutical Sciences, Fukuoka University, Japan. harashu@fukuoka-u.ac.jp
Abstract:
A rapid and sensitive HPLC method for the simultaneous determination of paraquat and diquat in human serum has been developed. After deproteinization of the serum with 10% trichloroacetic acid, the samples were separated on a reversed-phase column, and subsequently reduced to their radicals with alkaline sodium hydrosulfite solution. These radicals were monitored with a UV detector at 391 nm. This method permitted the reliable quantification of paraquat over linear ranges of 50 ng - 10 microg/ml and 100 ng - 10 microg/ml for diquat in human serum. The within- and between-day variations are lower than 2.3 and 2.2%, respectively. This technique was also utilized to determine the paraquat and diquat serum levels in a patient who had ingested herbicide (Prigrox L) containing paraquat and diquat.