The ear as an alternative site for a pulse oximeter finger clip sensor
1Department of Respiratory Therapy, St Joseph Hospital, Nashua NH 03060, USA. jhaynes@sjh-nh.org
Background:
Finger clip pulse oximetry sensors are commonly used to obtain functional oxygen saturation readings (S(pO2)), but these sensors may perform poorly if the digit is poorly perfused or there is excessive hand movement. I have increasingly witnessed clinicians obtaining S(pO2) readings by placing the finger clip sensor on the patient's ear when an S(pO2) reading cannot be obtained from a finger.
Objective:
Determine if reliable S(pO2) readings can be obtained from a finger clip sensor placed on the ear.
Methods:
This was a prospective study with patients undergoing pulmonary function testing. The calculated functional oxygen saturation values from arterial blood gas analysis (S(aO2)) were compared with S(pO2) readings from a finger clip sensor placed on a finger (finger S(pO2)) and on the upper portion of an ear (ear S(pO2)). S(pO2) data were included in the study only if (1) the pulse rate from finger S(pO2) and ear S(pO2) differed by < or = 5 beats/min and (2) the photoplethysmographic waveform was stable and acceptable.
Results:
Data were obtained from 30 adult white patients. The number of S(pO2) readings that differed from the S(aO2) values by > or = 3% were 1 (3.3%) finger S(pO2) reading and 24 (80%, 95% CI 61%-92%) ear S(pO2) readings (p < or = 0.001). Bland-Altman analysis showed better agreement between S(aO2) and finger S(pO2) (mean +/- 2 SD limits of agreement -2.35 to 2.35) than between S(aO2) and ear S(pO2) (limits of agreement -7.24 to -0.08) or finger S(pO2) and ear S(pO2) (limits of agreement -7.56 to -0.23).
Conclusion:
A pulse oximeter finger clip sensor placed on the ear does not provide clinically reliable S(pO2) readings.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse Assessment Sites
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: