Determination of 45 Pesticides in Avocado Varieties by the QuEChERS Method and Gas Chromatography-Tandem Mass Spectrometry
Published on: December 8, 2023
Pulsed large volume injection gas chromatography coupled with electron-capture negative ionization quadrupole mass
Yuli Zhao1, Limin Yang, Qiuquan Wang
1Department of Chemistry and the MOE Key Laboratory of Modern Analytical Sciences, College of Chemistry and Chemical Engineering, Xiamen University, Xiamen, China.
Abstract:
A pulsed large-volume injection gas chromatography coupled with electron-capture negative ionization quadrupole mass spectrometry (pLVI-GC/ECNI-qMS) was developed for the simultaneous determination of typical halogenated persistent organic pollutants (H-POPs). By monitoring the characteristic ions of large mass-to-charge ratio (m/z) for each of the H-POPs rather than the chlorine and/or bromine ions, this method avoided the possible interferences arising from the H-POPs themselves and from complex matrices encountered frequently in current GC/qMS methods; and allowed, on the other hand, the use of (13)C-labeled and perdeuterated analogues as internal standards for reliable quantification. pLVI up to 120 microL improved the instrumental detection limits down to pg-fg mL(-1), comparable to or lower than those obtained by the recognized GC/high-resolution MS methods reported so far. The H-POPs including 12 polybrominated diphenyl ethers, 1 polybrominated biphenyl, 10 polychlorinated biphenyls (PCBs), 4 hexachlorocyclohexane isomers, and hexachlorobenzene were involved in this study. The method developed demonstrated good linearity (r(2) = 0.9904-0.9999) within 0.5 to 50,000 pg mL(-1) for PCBs and 0.05 to 5000 pg mL(-1) for other H-POPs, and was satisfactory in terms of both repeatability (0.07%-2.2%) and reproducibility (2.1%-8.4%). It was validated by analyzing a NIST standard reference material SRM-1946 of Lake Superior fish tissue with low 0.01 to 63 pg g(-1) method detection limits, and successfully applied to the determination of the H-POPs in five reference materials of different matrices.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-II
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Alkyl Halide Fragmentation
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...