A new mechatronic assistance system for the neurosurgical operating theatre: implementation, assessment of accuracy
Jens Rachinger1, Klaus Bumm, Jochen Wurm
1Department of Neurosurgery, University of Erlangen-Nuremberg, Erlangen, Germany. jens.rachinger@medizin.uni-halle.de
Objective:
To introduce a new robotic system to the field of neurosurgery and report on a preliminary assessment of accuracy as well as on envisioned application concepts. Based on experience with another system (Evolution 1, URS Inc., Schwerin, Germany), technical advancements are discussed.
Material/Methods:
The basic module is an industrial 6 degrees of freedom robotic arm with a modified control element. The system combines frameless stereotaxy, robotics, and endoscopy. The robotic reproducibility error and the overall error were evaluated. For accuracy testing CT markers were placed on a cadaveric head and pinpointed with the robot's tool tip, both fully automated and telemanipulatory. Applicability in a clinical setting, user friendliness, safety and flexibility were assessed.
Results:
The new system is suitable for use in the neurosurgical operating theatre. Hard- and software are user-friendly and flexible. The mean reproducibility error was 0.052-0.062 mm, the mean overall error was 0.816 mm. The system is less cumbersome and much easier to use than the Evolution 1.
Conclusions:
With its user-friendly interface and reliable safety features, its high application accuracy and flexibility, the new system is a versatile robotic platform for various neurosurgical applications. Adaptations for different applications are currently being realized.