Compliance with alarm limits for pulse oximetry in very preterm infants
Luisa Clucas1, Lex W Doyle, Jennifer Dawson
1Department of Obstetrics and Gynaecology, University of Melbourne, Melbourne, Australia.
Insights
Compliance with hospital guidelines for pulse oximetry alarm limits in preterm infants on oxygen therapy was assessed. The lower alarm limit was frequently set correctly, but the upper limit was often set too high, especially in less vulnerable infants.
Area of Science:
- Neonatal Medicine
- Clinical Practice Guidelines
- Medical Device Monitoring
Background:
- Pulse oximetry is crucial for monitoring oxygenation in preterm infants.
- Hospital guidelines exist for setting alarm limits to ensure safe oxygen therapy.
- Compliance with these guidelines is essential for optimal patient care.
Purpose of the Study:
- To evaluate adherence to hospital guidelines for pulse oximetry alarm limits in preterm infants receiving oxygen.
- To identify factors influencing the correct setting of these alarm limits.
Main Methods:
- Prospective data collection on alarm limits in preterm infants (<32 weeks gestation or <1500g birth weight).
- Comparison of set alarm limits against hospital guidelines (target 88-92%, alarms at 85% and 94%).
- Analysis of factors associated with correct alarm limit setting.
Main Results:
- The lower alarm limit was correctly set in 91.1% of cases.
- The upper alarm limit was correctly set in only 23.3% of cases, frequently set too high (76.5%) or at 100% (23.8%).
- Correct upper alarm limit setting was associated with higher risk factors like lower birth weight and need for ventilation.
Conclusions:
- Current guidelines for upper pulse oximeter alarm limits in infants on oxygen therapy may be inconsistently followed.
- Compliance with lower alarm limits appears better than for upper limits.
- Higher-risk infants may have a greater likelihood of appropriately set upper alarm limits.
Objective:
The objective of this study was to determine the rate of compliance with hospital guidelines for alarm limits for pulse oximetry in preterm infants on oxygen therapy.
Methods:
All infants admitted to the nurseries at the Royal Women's Hospital, Melbourne, Australia, with gestational age <32 weeks or birth weight <1500 g between August 2005 and February 2006 were eligible for inclusion. Data on the alarm limits set for infants on oxygen therapy were collected prospectively. The target saturation range recommended in written hospital guidelines was 88% to 92%, with alarm limits set at 85% and 94%.
Results:
Data were prospectively collected for 144 subjects with mean (SD) gestational age 29.3 (2.4) weeks and birth weight 1226 (354) g; 1073 alarm limits were collected when infants were on oxygen. The lower alarm limit was set correctly 91.1% of the time. In contrast, the upper alarm limit was set correctly only 23.3% of the time: 76.5% of the time it was too high, and 23.8% of the time it was set at 100%. Infants with an upper alarm limit set correctly on a particular day had a significantly lower birth weight, gestational age, postmenstrual age, and postnatal age than infants who had the upper alarm limit set too high. Use of assisted ventilation, higher inspired oxygen concentrations, and more frequent changes in inspired oxygen concentration were all associated with improved odds of having an appropriately set upper alarm limit.
Conclusions:
This study suggests that current guidelines regarding the upper pulse oximeter alarm limit for infants receiving oxygen might be commonly exceeded, although compliance might be better for infants at higher risk of adverse outcomes. However, there might be less variation from guidelines for the lower alarm limit.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:27How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Measurement of Blood Pressure