Related Experiment Video
Updated: Jul 14, 2026

Voltage Biasing, Cyclic Voltammetry, & Electrical Impedance Spectroscopy for Neural Interfaces
Published on: February 24, 2012
Passive neutralization of myoelectric interference from neural recording tripoles
Ioannis Pachnis1, Andreas Demosthenous, Nick Donaldson
1Department of Electronic and Electrical Engineering, University College London, Torrington Place, London WC1E 7JE, UK.
Abstract:
In this paper, we present a simple passive technique for removing myoelectric interference in neural recording systems using tripolar electrodes. Imbalance is simply unavoidable with the conventional quasi-tripole (QT) amplifier and this technakshnique is based on a modified version of the QT, which sees the tripole as a bridge that can be balanced by adding impedance to one of the outer electrodes. We show that parallel resistance and capacitance is better than a series combination for use at all frequencies, and that with a tripole that was intentionally made imbalanced, by an amount that we measure as 3%, the interference can be reduced about 10-fold. It is important to null the interference at low frequencies, in the electromyography (EMG) band, to get the best improvement. Common-mode interference can also be reduced by appropriate trimming if necessary.