Related Experiment Video
Updated: Jul 14, 2026

Murine Endoscopy for In Vivo Multimodal Imaging of Carcinogenesis and Assessment of Intestinal Wound Healing and Inflammation
Published on: August 26, 2014
Video capsule endoscopy in patients with known or suspected small bowel stricture previously tested with the
Cristiano Spada1, Saumil K Shah, Maria Elena Riccioni
1Digestive Endoscopy Unit, Catholic University, Rome, Italy.
Goals:
To assess the safety of the Pillcam in patients with known or suspected radiologic stricture, previously tested for small bowel patency using the Given Patency capsule.
Background:
Intestinal stricture contributes a major contraindication to video capsule endoscopy (VCE), because of the risk of capsule retention.
Study:
Twenty-seven patients (16 female, mean age 44.2 y) with known or suspected intestinal stricture were enrolled prospectively. Twenty-four had Crohn's disease, 2 had adhesive syndrome and 1 had a suspected ischemic stricture. Patients underwent the Patency capsule test. In patients in whom the Patency capsule was excreted intact within 72 hours postingestion without occurrence of any adverse event, VCE was performed to assess the presence of strictures or other gastrointestinal pathologies. The following parameters were evaluated: transit time of Patency capsules and/or tags from ingestion to excretion, condition of the Patency capsule at excretion, transit time of the Pillcam capsule, the ability of Pillcam capsule to detect intestinal strictures and small bowel pathologies, any adverse events.
Results:
Twenty-five patients (92.6%) retrieved the Patency capsule in the stools. Six patients complained of abdominal pain, 4 of whom excreted a nonintact capsule. Hospitalization was required in 1 (4.3%) patient with Crohn's disease due to occlusive syndrome. Fifteen patients (65.3%) excreted an intact Patency capsule after a mean transit time of 25.6 hours without any adverse events. These 15 patients underwent the VCE successfully.
Conclusions:
Passage of an intact Patency capsule across a small bowel stricture provides direct evidence of functional patency of the gut lumen and allows a safe VCE. Intestinal strictures should not be considered an absolute contraindication for VCE.
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures V: ERCP
Patient...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Endoscopic Procedures II: Colonoscopy