Evaluation of finger and forehead pulse oximeters during mild hypothermic cardiopulmonary bypass
Ken Yamaura1, Kazuo Irita, Tadashi Kandabashi
1Department of Anesthesiology and Critical Care Medicine, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan.
Objective:
The purpose of this study was to examine and compare the four combination of pulse oximeters (POs) and monitoring sites, the Nihon Kohden BSS-9800 (N), the Masimo SET Radical (M), the Nellcor N550 D-25 (N-D) and the Nellcor N550 Max-Fast (N-MF) in patients with peripheral hypoperfusion.
Methods:
About 20 adult patients undergoing cardiac surgery using mild hypothermic cardiopulmonary bypass (CPB) were studied prospectively. PO sensors were applied on fingers in N, M and N-D, while on the forehead in N-MF.
Results:
PO failure was defined as failure to show no SpO2 value or incorrect SpO2 values. PO failure occurred in 12 patients with N, ten patients with M, four patients with N-D and ten patients with N-MF, respectively (p < 0.05 N-D vs. N, M, N-MF). The duration of PO failure was 19 +/- 30% of aortic cross-clamping with N, 29 +/- 33% with M, 10 +/- 26% with N-D and 43 +/- 57% with N-MF, respectively (p < 0.05 N-D vs. M and N-MF).
Conclusions:
The results suggested that N-D is most useful among four combinations of POs and monitoring sites tested in this study for monitoring SpO2 during hypoperfusion. The superiority of N-MF during hypoperfusion was not evident in the present study.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation