Thoracic aortic stent-graft devices: problems, failure modes, and applicability.
Christoph A Nienaber1, Stephan Kische, Hüseyin Ince
1Division of Cardiology, University Hospital Rostock, Rostock School of Medicine, Rostock, Germany. christoph.nienaber@med.uni-rostock.de
Seminars in Vascular Surgery
|June 21, 2007
Summary
Thoracic endografts offer less invasive treatment for aortic pathologies, but design limitations hinder optimal outcomes. Further advancements are needed for better conformability and fixation in the aortic arch.
Area of Science:
- Cardiovascular Surgery
- Medical Device Engineering
- Vascular Medicine
Background:
- Treatment for descending thoracic aorta pathologies remains debated, with open surgery being complex.
- Endovascular devices offer a less invasive alternative, enabling nonsurgical access and maintaining aortic blood flow without clamping.
- Current short- and mid-term outcomes for endografting thoracic aneurysms and type B aortic dissections show promise, with reduced morbidity and mortality compared to open surgery.
Purpose of the Study:
- To provide an overview of thoracic aortic stent-graft devices.
- To highlight existing problems and failure modes associated with current thoracic endografts.
- To discuss potential improvements in thoracic endograft design and delivery systems.
Main Methods:
- Review of current literature on thoracic aortic stent-graft devices.
- Analysis of challenges in endovascular repair of thoracic aorta pathologies.
- Discussion of limitations in endograft design, delivery systems, and fixation.
Main Results:
- Current endograft designs face challenges with rigidity, delivery system size, and conformability to the aortic arch anatomy.
- Nonconformity can lead to graft instability, endoleak, and procedural failure.
- Delivery systems can be traumatizing and inflexible, often requiring surgical vessel exposure.
Conclusions:
- Despite encouraging clinical outcomes, thoracic endograft technology has not advanced sufficiently to meet clinical demands.
- Significant improvements are required in miniaturization, conformability, and disease-specific device creation for thoracic aortic repair.